Patient Monitor

Multi-parameter Patient Monitor YSPM80C

Model

YSPM80C

Brand

YSENMED

Product Description

Technical Specification
NIBP Method Oscillometry
Measure mode Manual, Auto, STAT
Measure interval in auto mode 1~5/10/15/30/60/90/120/180/240/480min
Measure period in STAT mode 5min
Pulse rate range 40 ~ 240bpm
Measure & alarm range: Adult
SYS 40 ~ 280mmHg
DIA 10 ~ 220mmHg
MEAN 20 ~ 240mmHg
Pediatric
SYS 40 ~ 220mmHg
DIA 10 ~ 160mmHg
MEAN 20 ~ 170mmHg
Neonatal
SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20 ~ 110mmHg
Static pressure measurement
Range 0 ~ 300mmHg
Accuracy ± 3mmHg
Resolution / Accuracy 1mmHg / Max. Mean Error: ± 5mmHg / Max. Standard Deviation: 8mmHg
Overpressure protection Adult 300mmHg / Pediatric 240mmHg / Neonatal 150mmHg
SPO2 Measurement range 0 ~ 100%
Resolution 1%
Accuracy 70% ~ 100% ± 2% / 40% ~ 69% ± 3%
Environment Operating environment
Temperature 0 ~ 40 ℃
Humidity / Pressure 15% ~ 80% / 86.0 kPa ~ 106.0kPa
Transport and Storage environment
Temperature -20 ~ 55 ℃
Humidity / Pressure 10% ~ 90% / 86.0 kPa ~ 106.0kPa