UV Spectrophotometer
UV Spectrophotometer/Visible Spectrophotometer YSTE-UV1000
Model 4194_92ff8f-07> |
YSTE-UV1000 4194_1b5ce2-51> |
Brand 4194_a35474-11> |
YSENMED 4194_eb7607-79> |
Product Description
Model: YSTE-UV1000
Features:
- Saves up to 200 groups of data and 200 standard curves.
- Data is retained even after sudden power failure.
- Auto setting of wavelengths to calibrate the system via arrow keys.
- Wavelength range: 190 to 1100 nm.
- Automatic switching between Tungsten & Deuterium lamps.
- Fully equipped for standalone operation.
- Large sample compartment supports 5–100 mm path length cuvettes with optional holders.
- Optional ‘Wave Professional’ software enables PC control, spectrum scan, time scan (kinetics), multi-wavelength scanning, and nucleic acid/protein measurement.
- Wide range of compatible accessories available.
Description:
User-Friendly Operation and Rich LCD Display: Features a large 128×64 backlit LCD screen with adjustable brightness for displaying comprehensive data and graphical formats.
Onboard Data Storage: Capable of storing up to 200 test results and 200 standard curves in RAM, with recovery after power interruptions.
Auto Wavelength Calibration: Allows automatic wavelength calibration to reduce manual errors.
Technical Specifications:
Spectral Bandwidth | 4.0 nm |
Wavelength Accuracy | ±2 nm |
Wavelength Repeatability | ≤1 nm |
Photometric Range | -0.3–3.0 A, 0–200%T |
Photometric Accuracy | ±0.003 A @ 0.5 A ±0.006 A @ 1.0 A ±0.5%T @ 100%T |
Photometric Repeatability | ≤0.0015 A @ 0.5 A ≤0.003 A @ 1.0 A ≤0.25%T @ 100%T |
Stability | ≤0.002 A/h @ 500 nm (after 2 hours preheat) |
Stray Light | ≤0.2%T @ 220 & 360 nm |
Sample Compartment | 4-position, 10 mm pathlength cuvette |
Display | LCD (128 × 64) |
Output | USB Port & Parallel Port (Printer) |
Power | 110/220V AC, 50/60Hz, 120W |
Other Attributes:
Model Number | YSTE-UV1000 |
Spectral Bandwidth | 4.0 nm |
Optical System | Single Beam |
Light Source | Tungsten lamp & Deuterium lamp |
Wavelength Range | 200–1000 nm |
Wavelength Accuracy | ±2 nm |
Photometric Range | -0.3–3.0 A, 0–200%T |
Stray Light | ≤0.2%T @ 220 & 360 nm |
Power Requirements | 110/220V AC, 50/60Hz, 120W |
Dimension (L×W×H) | 490×360×210 mm |
Net Weight | 12 kg |