Patient Monitor

15 Inch Medical Modular Patient Monitor YSPM-G3LM

Model

YSPM-G3LM

Brand

YSENMED

Product Description

Features

• Dual plug-ins: EtCO2, 2lBP, 4lBP, CO, AG can be optional; modular monitor is more convenient for detection and more economical;

• Data: Trend data storage for up to 360 hours; printing for data, waveform, and trend diagrams;

• User Interface: 15″ color TFT LCD screen, available in 15 languages;

• Versatility: Suitable for adults, pediatrics, and neonates; widely applicable in physician’s offices, OR/OT, ICU, CCU, NiCU, clinics, etc

• Communication: WAN communication function to network with Central Monitoring System, supporting both wired and wireless networking and compatibility with HL7 systems;

• Accuracy Assurance: “Auto-current drift balanced system” ensure accuracy, along with unique patents for impedance RESP and ECG waveform stability during electrosurgical procedures;

• Advanced Functions: includes ST-segment analysis, bed-to-bed view, venipuncture support, pacemaker monitoring, arrhythmia analysis and 5 calculations (Hemodynamic, Drug Dose, Oxygenation, Renal Function, Ventilation);

• Specialized Tests: Offers unique nasal tube respiration testing suitable for patients with burns, upper airway obstructions, and newborns.


Technical Specification
ECG Input 3/5/10 wires ECG cable
Lead section I, ll, lll, aVR, aVL, aVF, V1-V6
Gain selection x1/4, x1/2, x1, x2, x4 and automatic
Sweep speed 3.125, 6.25, 12.5, 25, 50mm/s
Heart rate range 15~350bpm
Calibration ±1mv
Accuracy ±1bpm or +1% (select the bigger data)
SPO2 Display type Waveform, Data
Measurement range 0~100%
Accuracy ±3% (between 70%~100%)
SpO2 resolution 1%
Pulse rate range 25~250bpm
Accuracy ±1bpm or ±2% (select the bigger data)
PR Resolution 1bpm
PI Measurement range 0.00~20.0%
Accuracy Not defined
Resolution 0.00~9.99%: 0.01% / 10.0~20.0%: 0.1%
Respiration Measurement range 0~120rpm
Accuracy ±1rpm or ±5% (select the bigger data)
Resolution 1rpm
EtCO2 – Plug & Play (Option) Measurement type Main/Side stream
Measurement range 0%~19.7%, 0~150mmHg, 0~20kpa
Accuracy 0~40mmHg ±2mmHg / 41~70mmHg ±5% of reading / 71~100mmHg ±8% of reading / 101~150mmHg ±10% of reading
2-Temperature (Rectal & Surface) Number of channel 2 channel
Measurement range 0°C ~50°C
Accuracy ±0.2°C
Resolution 0.1°C
Display T1, T2, ∆T / Unit: °C/°F selection
2-IBP/4-IBP (Option) Measurement range -10~300mmHg
Number of channel 2 or 4 channel
Accuracy ±1mmHg or ±2% (select the bigger data)
IBP Pressure Tag ART, CVP, RVP, LAP, RAP, PAP, LCP, LVP, AO, FAR, BAP, VAP, LV, PA, P1, P2, P3, P4
NIBP Philosophy Oscillometric method
Measurement parameter Systolic, Diastolic, Mean
Measurement method Manual, Auto, Continuous
Measuring range SYS: 30-270mmHg / DIA: 10-200mmHg / MAP: 20-220mmHg
Accuracy Max average error: ±5mmHg / Max standard deviation: 8mmHg / Resolution: 1mmHg
Cardiac Output (Option) Measuring Method Thermodilution
Measurement range C.O.: 0.1~20.0L/min / TB: 20.0°C~45.0°C / Tl: 0.0°C~30.0°C
Accuracy C.O.: ±5% or ±0.1L/min (select the bigger data) / TB: ±0.2°C / Tl: ±0.2°C
Resolution CO: 0.1L/min / TB: 0.1°C / TI: 0.1°C
Anesthesia Gas Analysis (Option) Method Infrared Absorption
Gas Sorts CO2, N2O, Des, lSO, Enf, Sev, Hal and optional O2
Measurement range CO2: 0~13.6%; N2O: 0~100%; O2: 0~100%; Hal: 0~8.5%; Iso: 0~8.5%; Enf: 0~10%; Sev: 0~10%; Des: 0~20%
RR (Respiration rate) 0~100rpm
Data input Fi and Et values / Agent mixture detection
Standard Configuration ECG, NIBP, SPO2, RESP, 2TEMP, HR/PR, Rechargeable battery